Electroanalytical determination of soluble Mn(II) species at high concentration levels

https://doi.org/10.1016/j.electacta.2017.04.066Get rights and content

Abstract

Glassy carbon electrodes have been successfully employed for the determination, by differential pulse voltammetry, of Mn(II) ions dissolved in aqueous solutions. In particular, a simple and fast procedure also suitable for on-line or at-line process control has been developed. Statistical analysis of the results obtained reveals that the procedure can be adopted in the range 0.59–57.30 mM Mn(II). The electrochemical responses are repeatable and reproducible. Repeatability is testified by the lack of variation of peak current values calculated from 20 subsequent scans carried out at the maximum explored concentration (relative standard deviation <1%). Reproducibility of the responses is supported by the data from the responses on three different electrodes. The proposed procedure does not require any pre-concentration of Mn species at the electrode surface or de-aeration of the solution. By adopting the developed procedure for the analysis, the GC electrode demonstrates to be suitable also for application in real matrices, namely solutions from spent battery recycling; in particular the results from the present electrochemical method are not significantly different from those obtained through inductively plasma coupled mass spectrometry. Interference from other metal species, such as Zn(II) and Fe(III), is negligible. This result is particularly meaningful, since the experimental conditions chosen, implying particularly high concentrations of heavy metals, are representative of those adopted in recovery and recycle processes of Mn species from batteries.

Introduction

Dissolved Mn(II) ions are ubiquitous species that are present, for instance, in drinking water, foodstuff, plants, etc. [1], [2]. Reasonably, most analytical methods developed so far focus the attention on the determination of traces of Mn species. In any case, the most widely diffused protocols for the quantitative evaluation of these species require the adoption of off-line procedures, implying sampling, transfer of samples to the laboratory equipped for the purpose and pre-treatment of the sample before the analysis. Methods, such as those based on atomic absorption or on emission spectroscopies, are well established [1]. Among few spectrophotometric methods in the UV–vis region for analyses in solution [3], the oxidation to Mn(VII) or the addition of formaldoxime or 1-(2-pyridylazo)-2-naphthol as complexing chromophores deserve citation. Unfortunately, a number of limitations have been reported in the cited Refs. [1], [3]; as an example, methods based on redox reactions suffer from the interference of other oxidisable species whilst those exploiting the formation of coloured complexes are adversely affected by the presence of heavy metals different from Mn.

Electroanalytical methods possess a number of advantages over other analytical techniques, such as simplicity, low cost, and possibility to operate in turbid solutions, which is quite a common situation encountered in real matrices, as it is the case examined here [1]. Once proper electrode systems and relevant procedures are developed, they are suitable for on-line and at-line analyses, i.e., using portable instrumentation, without the necessity to bring a sample to external laboratories [4]. At the moment, only few literature reports deal with the electrochemical determination of Mn(II) ions (see, for instance, Ref. [5] and a survey of the most significant papers reported in the Supplementary Materials). Many approaches make use of Hg-based electrodes; it is however well known that, despite the effectiveness of this electrode material in the determination of heavy metals at the trace level, it is progressively phased out from the laboratories and industrial plants, due to its toxicity. This forces to identify alternative electrode materials for effective electrochemical determination of a variety of chemicals; this is particularly urgent for heavy metal species.

Apart from trace analysis, a basically different, though comparably important, issue regards the quantification of Mn species present at high concentrations. These are conditions often encountered in a variety of industrial frames. In order to reduce the cost derived from the use of big instrumentation and the time required by transport and complex treatment of the sample, the use of sensing systems is essential. In situ continuous monitoring of Mn species is necessary in many cases, such as recycling of batteries, metal electrowinning, and production of fertilizers [6], [7], [8], [9], [10].

In principle, high dilution ratios could allow the use of the previously cited literature methods. However, such an approach is impractical for an effective at-line or on-line process control, due to the high dilution volumes that may be required. Additional points prevent from an effective application of established electroanalytical procedures: i) the necessity to remove dissolved oxygen by bubbling an inert gas, when anodic stripping analyses are performed [11], [12]; ii) the low mechanical robustness of many electrode modifications, e.g. carbon pastes, potentially effective to the purpose [13], [14]; iii) the high pH values required by the methods, that may lead to precipitation of Mn hydroxides at high Mn concentration level [15], [16].

In the present paper, a procedure suitable for the determination of high concentrations of Mn(II) ions (0.59–57.30 mM) is presented. A simple glassy carbon (GC) electrode is used, once the conditions under which to obtain repeatable responses are set. It is worth noticing that this electrode material has been employed very rarely in the determination of Mn(II) ions, limitedly to the determination of traces (see, e.g., [17], [18], [19] and Table S-1 in Supplementary Materials). The method proposed here does not require any accumulation step. In addition, regeneration of the electrode surface between subsequent measurements, such as mechanical polishing, is unnecessary and even deaeration of the solution is not required: samples can be analysed without any specific pre-treatment.

Moreover, it should be emphasised that no literature report considers the concentration interval to which the present paper is devoted. Noteworthy, the electroanalytical determination of species, in particular of metal ions, at high concentrations requires to overcome a number of drawbacks, due to possible enhancement of the effect of poisoning adsorptions or enhancement of undesired side reactions, etc.

Finally, the interference of different heavy metals and of oxygen is negligible, at variance with a number of literature reports. The effectiveness of the analytical procedure has been also checked by us, operating on an untreated and particularly challenging real matrix, namely the leachate from the recycling process of spent alkaline batteries [9]. The results from the present method are not significantly different from those obtained through inductively plasma coupled mass spectrometry.

Section snippets

Chemicals and sample preparation

Chemicals, namely MnSO4·H2O, ZnSO4·7H2O, FeCl3·6H2O, CuCl2·2H2O, NiSO4·6H2O, PbCl2 and 96% w/w H2SO4, were from Sigma. In addition, Cd(II), Hg(II) and Co(II) standard solutions (1000 ppm from Peak Performance) have been employed. All solutions were prepared using ultrapure water (18  cm resistivity, MILLIQ plus by MILLIPORE). A multistandard from Inorganic Venture (IV-ICPMS-71A) has been employed for the calibration of the ICP instrument.

MnO2-carbon powder mixture from commercial spent alkaline

Voltammetric response

The usual voltammetric behaviour of Mn(II) ions in acidic solutions is reported in the literature [23], [24]. In most cases the studies were focussed on the formation of thick and stable MnO2 films on metal electrode surfaces, in the frame of the realisation of capacitors [25], [26]. The voltammetric behaviour of high concentrations of Mn(II) (0.59–57.30 mM) has been studied by us in 1 M H2SO4. The choice of the nature and concentration of the solvent medium was made in order to fit the typical

Conclusions

The necessity to define best conditions for amperometric determination of analytes of specific interest in practical frames encourages to look for novel procedures and electrode systems. On the other hand, while our recent activity has been addressed mainly to novel, unusual electrode materials, either mono- or bimetallic in formulation, based on Ti and Ti-Au bimetallic systems [33], [36], the traditional GC electrodes have confirmed to deserve much attention, once suitable measurement

Acknowledgements

This work was supported by the Italian Ministry of University and Scientific Research (grant RBFR13NZH9).